Quantum calibrated magnetic force microscopy
arXiv:2103.13785 · doi:10.1103/PhysRevB.104.214427
Abstract
We report the quantum calibration of a magnetic force microscope (MFM) by measuring the two-dimensional magnetic stray field distribution of the MFM tip using a single nitrogen vacancy (NV) center in diamond. From the measured stray field distribution and the mechanical properties of the cantilever a calibration function is derived allowing to convert MFM images to quantum calibrated stray field maps. This novel approach overcomes limitations of prior MFM calibration schemes and allows quantum calibrated nanoscale stray field measurements in a field range inaccessible to scanning NV magnetometry. Quantum calibrated measurements of a stray field reference sample allow its use as a transfer standard, opening the road towards fast and easily accessible quantum traceable calibrations of virtually any MFM.
arXiv comment31 Pages, 5 Figures
References in corpus (8)
- Quantum sensing
- The nitrogen-vacancy colour centre in diamond
- Probing condensed matter physics with magnetometry based on nitrogen-vacancy centres in diamond
- Field-Dependent Size and Shape of Single Magnetic Skyrmions
- Individual skyrmion manipulation by local magnetic field gradients
- Analytic theory for the switch from Bloch to Néel domain wall in nanowires with perpendicular anisotropy
- Theoretical analysis of magnetic force microscopy contrast in multidomain states of magnetic superlattices with perpendicular anisotropy
- Modal frustration and periodicity breaking in artificial spin ice