Systematic uncertainties in integrated luminosity measurement at CEPC

arXiv:2010.15061 · doi:10.1088/1748-0221/17/09/P09014

Abstract

The very forward region is one of the most challenging regions to instrument at a future collider. At CEPC, machine-detector interface includes, among others, a calorimeter dedicated for precision measurement of the integrated luminosity at a per mill level or better. Here we review a feasibility of such precision, from the point of view systematic effects arising from luminometer mechanical precision and positioning, beam-related requirements and physics background from two-photon processes. The impact of the beam energy spread and its uncertainty on the integrated luminosity precision is also discussed, as well as the achievable beam energy spread precision with the post-CDR CEPC beams.

12 pages, 10 figures, 4 tables