Active Optical Table Tilt Stabilization
arXiv:2003.03404 · doi:10.1063/5.0006916
Abstract
We show that a simple modification to an optical table with pneumatic vibration isolation can be used to actively reduce the long term drift in the tilt of the table by nearly a factor of 1000. Without active stabilization, we measure a root-mean-square (RMS) tilt variation of \SI{270}{\upmu rad} over three days. The active stabilization can be used to limit the tilt to \SI{0.35}{\upmu rad} RMS over the same time period. This technique can be used to minimize drift in tilt-sensitive experiments.
arXiv comment3 pages, 4 figures
References in corpus (4)
- Sideband Cooling Micromechanical Motion to the Quantum Ground State
- Zeptonewton force sensing with nanospheres in an optical lattice
- Test of the isotropy of the speed of light using a continuously rotating optical resonator
- Cooling the Motion of a Silica Microsphere in a Magneto-Gravitational Trap in Ultra-High Vacuum