Characterization of High-Purity Germanium (Ge) Crystals for Developing Novel Ge Detectors
arXiv:2002.07706 · doi:10.1088/1748-0221/15/10/T10010
Abstract
High-purity germanium (HPGe) crystals are required to be well-characterized before being fabricated into Ge detectors. The characterization of HPGe crystals is often performed with the Hall Effect system, which measures the carrier concentration, the Hall mobility, and the electrical resistivity. The reported values have a strong dependence on the size of the ohmic contacts and the geometry of the samples used in conducting the Hall Effect measurements. We conduct a systematic study using four samples cut from the same location in a HPGe crystal made into different sized ohmic contacts or different geometries to study the variation of the measured parameters from the Hall Effect system. The results are compared to the C-V measurements provided by the Ge detector made from the same crystal. We report the systematic errors involved with the Hall Effect system and find a reliable technique that minimizes the systematic error to be only a few percent from the Hall Effect measurements.
arXiv comment11 page, 4 figures, and 6 Tables
References in corpus (10)
- Results from a Search for Light-Mass Dark Matter with a P-type Point Contact Germanium Detector
- Evidence for Neutrinoless Double Beta Decay
- First Dark Matter Constraints from a SuperCDMS Single-Charge Sensitive Detector
- Search for Zero-Neutrino Double Beta Decay in 76Ge with the Majorana Demonstrator
- Limits on Light Weakly Interacting Massive Particles from the First 102.8 kg day Data of the CDEX-10 Experiment
- Probing Majorana neutrinos with double- decay
- Improved EDELWEISS-III sensitivity for low-mass WIMPs using a profile likelihood approach
- Search for Light Weakly-Interacting-Massive-Particle Dark Matter by Annual Modulation Analysis with a Point-Contact Germanium Detector at the China Jinping Underground Laboratory
- Direct Detection of MeV-Scale Dark Matter Utilizing Germanium Internal Amplification for the Charge Created by the Ionization of Impurities
- Fabrication and Characterization of High-Purity Germanium Detectors with Amorphous Germanium Contacts