Flux compensation for SQUID-detected Magnetic Resonance Force Microscopy
arXiv:1810.12571 · doi:10.1016/j.cryogenics.2019.01.001
Abstract
One of the major challenges in performing SQUID-detected Magnetic Resonance Force Microscopy (MRFM) at milliKelvin temperatures is the crosstalk between the radiofrequency (RF) pulses used for the spin manipulation and the SQUID-based detection mechanism. Here we present an approach based on balancing the flux crosstalk using an on-chip feedback coil coupled to the SQUID. This approach does not require any additional components near the location of the sample, and can therefore be applied to any SQUID-based detection scheme to cancel predictable RF interference. We demonstrate the effectiveness of our approach by showing that we can almost completely negate flux crosstalk with an amplitude of up to several . This technical achievement allows for complicated magnetic resonance protocols to be performed at temperatures below 50 mK.