A compact ultrahigh vacuum scanning tunneling microscope with dilution refrigeration
arXiv:1806.03171 · doi:10.1063/1.5043636
Abstract
We have designed and built a scanning tunneling microscope (STM) setup for operation at millikelvin temperatures in ultra high vacuum. A compact cryostat with an integrated dilution refrigerator has been built, that allows measurements at a base temperature of 25 mK in magnetic field up to 7.5 T with low mechanical and electronic noise. The cryostat is not larger than conventional helium bath cryostats (23 and 13 liters of nitrogen and helium respectively), so that the setup does not require a large experimental hall and fits easily into a standard lab space. Mechanical vibrations with running dilution circulation were kept below 300 fm/ by mechanically decoupling the STM from the cryostat and the pumping system. All electronic input lines were low-pass filtered, reducing the electronic temperature to below 100 mK, as deduced from the quasiparticle peaks of superconducting aluminium. The microscope is optically accessible in the parked position, making sample and tip exchange fast and user-friendly. For measurement the STM is lowered 60 mm down so that the sample ends in the middle of a wet superconducting magnetic coil.
arXiv comment9 pages, 1 table, 9 figures. Submitted to Rev. Sci. Instr
References in corpus (7)
- Imaging the Effects of Individual Zinc Impurity Atoms on Superconductivity in Bi2Sr2CaCu2O8+delta
- Magnetic Anisotropy and Magnetization Dynamics of Individual Atoms and Clusters of Fe and Co on Pt(111)
- The energy resolution function of a tunnel junction
- Design and performance of an ultra-high vacuum scanning tunneling microscope operating at dilution refrigerator temperatures and high magnetic fields
- A scanning tunneling microscope for spectroscopic imaging below 90 mK in magnetic fields up to 17.5 T
- Design and performance of an ultra-high vacuum spin-polarized scanning tunneling microscope operating at 30 mK and in a vector magnetic field
- A 30 mK, 13.5 T scanning tunneling microscope with two independent tips
Cited by in corpus (15)
- Designer quantum states of matter created atom-by-atom
- A modular ultra-high vacuum millikelvin scanning tunneling microscope
- Microscopic charging and in-gap states in superconducting granular aluminum
- Unconventional pairing in single FeSe layers
- A millikelvin scanning tunneling microscope in ultra-high vacuum with adiabatic demagnetization refrigeration
- How cold is the junction of a millikelvin scanning tunnelling microscope?
- Sparse Sampling for Fast Quasiparticle Interference Mapping
- Identification of multiple-flux-quanta vortices by core states in the two-band superconductor Pb
- Proximity-Induced Superconductivity in Atomically Precise Nanographene
- Magnetic bound states embedded in tantalum superconducting thin films
- Performance benchmarking of an ultra-low vibration laboratory to host a commercial millikelvin scanning tunnelling microscope
- Friedel Oscillations and superconducting-gap enhancement by impurity scattering
- Removing background and estimating a unit height of atomic steps from a scanning probe microscopy image using a statistical model
- Visualization of defect-induced interband proximity effect at the nanoscale
- A scanning tunneling microscope capable of electron spin resonance and pump-probe spectroscopy at mK temperature and in vector magnetic field