Huddle test measurement of a near Johnson noise limited geophone
arXiv:1711.05439 · doi:10.1063/1.5000592
Abstract
In this paper the sensor noise of two geophone configurations (L-22D and L-4C geophones from Sercel with custom built amplifiers) was measured by performing two huddle tests. It is shown that the accuracy of the results can be significantly improved by performing the huddle test in a seismically quiet environment and by using a large number of reference sensors to remove the seismic foreground signal from the data. Using these two techniques, the measured sensor noise of the two geophone configurations matched calculated predictions remarkably well in the bandwidth of interest (0.01 Hz to 100 Hz). Low noise operational amplifiers OPA188 were utilized to amplify the L-4C geophone to give a sensor that was characterized to be near Johnson noise limited in the bandwidth of interest with a noise value of at 1 Hz.
arXiv comment7 pages, 2 supplementary materials, 8 figures plus 2 supplementary material figures, accepted for publication in Review of Scientific Instruments
References in corpus (5)
- Advanced Virgo: a 2nd generation interferometric gravitational wave detector
- Seismic isolation of Advanced LIGO: Review of strategy, instrumentation and performance
- Automatic cross-talk removal from multi-channel data
- The 10m AEI prototype facility A brief overview
- Passive-performance, analysis, and upgrades of a 1-ton seismic attenuation system
Cited by in corpus (8)
- High sensitivity accelerometry with a feedback-cooled magnetically levitated microsphere
- Improving the Robustness of the Advanced LIGO Detectors to Earthquakes
- A fifty-fold improvement of thermal noise limited inertial sensitivity by operating at cryogenic temperatures
- High Precision Inertial Sensors on a One Inch Diameter Optic
- A vertical inertial sensor with interferometric readout
- Designing Gram-Scale Resonators for Precision Inertial Sensors
- On the Development of an RFSoC-Based Ultra-Fast Phasemeter With GHz Bandwidth
- Interferometric sensing of a commercial geophone