Convergent-Beam EMCD: Benefits, Pitfalls, and Applications
arXiv:1707.01579 · doi:10.1093/jmicro/dfx129
Abstract
Energy-loss magnetic chiral dichroism (EMCD) is a versatile method for studying magnetic properties on the nanoscale. However, the classical EMCD technique is notorious for its low signal to noise ratio (SNR). Here, we study the theoretical possibilities of using a convergent beam for EMCD. In particular, we study the influence of detector positioning as well as convergence and collection angles on the detectable EMCD signal. In addition, we analyze the expected SNR and give guidelines for achieving optimal EMCD results.