Summing coincidence in rare event gamma-ray measurements under an ultra-low background environment
arXiv:1705.04316 · doi:10.1016/j.nima.2017.09.043
Abstract
A Monte Carlo method based on the GEANT4 toolkit has been developed to correct the full-energy peak (FEP) efficiencies of a high purity germanium (HPGe) detector equipped with a low background shielding system, and moreover evaluated using summing peaks in a numerical way. It is found that the FEP efficiencies of Co, Ba and Eu can be improved up to 18\% by taking the calculated true summing \mbox{coincidence} factors (TSCFs) correction into account. Counts of summing coincidence peaks in the spectrum of Eu can be well reproduced using the corrected efficiency curve within an accuracy of 3\%.
arXiv commentreplace with the accepted version