Online Rn removal by cryogenic distillation in the XENON100 experiment
arXiv:1702.06942 · doi:10.1140/epjc/s10052-017-4902-x
Abstract
We describe the purification of xenon from traces of the radioactive noble gas radon using a cryogenic distillation column. The distillation column is integrated into the gas purification loop of the XENON100 detector for online radon removal. This enabled us to significantly reduce the constant Rn background originating from radon emanation. After inserting an auxiliary Rn emanation source in the gas loop, we determined a radon reduction factor of R > 27 (95% C.L.) for the distillation column by monitoring the Rn activity concentration inside the XENON100 detector.