High accuracy position response calibration method for a micro-channel plate ion detector
arXiv:1605.08686 · doi:10.1016/j.nima.2016.08.024
Abstract
We have developed a position response calibration method for a micro-channel plate (MCP) detector with a delay-line anode position readout scheme. Using an {\em in situ} calibration mask, an accuracy of 8~m and a resolution of 85~m (FWHM) have been achieved for MeV-scale particles and ions with energies of 10~keV. At this level of accuracy, the difference between the MCP position responses to high-energy particles and low-energy ions is significant. The improved performance of the MCP detector can find applications in many fields of AMO and nuclear physics. In our case, it helps reducing systematic uncertainties in a high-precision nuclear -decay experiment.