Co-design of an in-line holographic microscope with enhanced axial resolution: selective filtering digital holography
arXiv:1601.02940 · doi:10.1364/JOSAA.33.000107
Abstract
Common-path digital in-line holography is considered as a valuable 3D diagnostic techniques for a wide range of applications. This configuration is cost effective and relatively immune to variation in the experimental environment. Nevertheless, due to its common-path geometry, the signal to noise-ratio of the acquired hologram is weak as most of the detector (i.e. CCD/CMOS sensor) dynamics is occupied by the reference field signal, whose energy is orders of magnitude higher than the field scattered by the imaged object. As it is intrinsically impossible to modify the ratio of energy of reference to the object field, we propose a co-design approach (Optics/Data Processing) to tackle this issue. The reference to object field ratio is adjusted by adding a 4-f device to a conventional in-line holographic setup, making it possible to reduce the weight of the reference field while keeping the object field almost constant. Theoretical analysis of the Cràmer-Rao lower bounds of the corresponding imaging model illustrate the advantages of this approach. These lower bounds can be asymptotically reached using a parametric inverse problems reconstruction. This implementation results in a 60 % gain in axial localization accuracy (for of 100 m diameter spherical objects) compared to a classical in-line holography setup .
References in corpus (4)
- (DH) Noise and Signal scaling factors in Digital Holography in week illumination: relationship with Shot Noise
- 3D tracking the Brownian motion of colloidal particles using digital holographic microscopy and joint reconstruction
- Vibration of low amplitude imaged in amplitude and phase by sideband versus carrier correlation digital holography
- Singular value demodulation of phase-shifted holograms