Simultaneous and accurate measurement of the dielectric constant at many frequencies spanning a wide range
arXiv:1503.07032 · doi:10.1063/1.4916260
Abstract
We present an innovative technique which allows the simultaneous measurement of the dielectric constant of a material at many frequencies, spanning a four orders of magnitude range chosen between 10 --2 Hz and 10 4 Hz. The sensitivity and accuracy are comparable to those obtained using standard single frequency techniques. The technique is based on three new and simple features: a) the precise real time correction of the amplication of a current amplier; b) the specic shape of the excitation signal and its frequency spectrum; and c) the precise synchronization between the generation of the excitation signal and the acquisition of the dielectric response signal. This technique is useful in the case of relatively fast dynamical measurements when the knowledge of the time evolution of the dielectric constant is needed.
References in corpus (8)
- Energy flow, partial equilibration and effective temperatures in systems with slow dynamics
- Fluctuation-dissipation ratio of a spin glass in the aging regime
- Evidence of growing spatial correlations at the glass transition from nonlinear response experiments
- Experimental verification of a modified fluctuation-dissipation relation for a micron-sized particle in a non-equilibrium steady state
- Fluctuation-Dissipation-Theorem violation during the formation of a colloidal-glass
- Fluctuation dissipation ratio in an aging Lennard-Jones glass
- Fluctuation-dissipation relation in a sheared fluid
- Intermittent origin of the large violations of the fluctuation dissipation relations in an aging polymer glass