An Efficient, Movable Single-Particle Detector for Use in Cryogenic Ultra-High Vacuum Environments
arXiv:1411.4530 · doi:10.1063/1.4907352
Abstract
A compact, highly efficient single-particle counting detector for ions of keV/u kinetic energy, movable by a long-stroke mechanical translation stage, has been developed at the Max-Planck-Institut für Kernphysik (Max Planck Institute for Nuclear Physics, MPIK). Both, detector and translation mechanics, can operate at ambient temperatures down to 10 K and consist fully of ultra-high vacuum (UHV) compatible, high-temperature bakeable and non-magnetic materials. The set-up is designed to meet the technical demands of MPIK's Cryogenic Storage Ring (CSR). We present a series of functional tests that demonstrate full suitability for this application and characterise the set-up with regard to its particle detection efficiency.
arXiv comment12 pages, 9 figures, version accepted for publication in Review of Scientific Instruments
References in corpus (1)
Cited by in corpus (6)
- The Cryogenic Storage Ring CSR
- Photoionization of ions with synchrotron radiation: From ions in space to atoms in cages
- Dielectronic recombination of lanthanide and low ionization state tungsten ions: W - W
- Recombination of open-f-shell tungsten ions
- Single-particle detection of products from atomic and molecular reactions in a cryogenic ion storage ring
- Unimolecular processes in diatomic carbon anions at high rotational excitation