Measurement of intrinsic radioactive backgrounds from Cs-137 and U/Th chains in CsI(Tl) crystals
arXiv:1408.0590 · doi:10.1088/1674-1137/39/4/046002
Abstract
The inorganic CsI(Tl) crystal scintillator is a candidate served as an anti-compton detector for China Dark matter EXperiment (CDEX). Studying the intrinsic radiopurity of CsI(Tl) crystal is an issue of major importance. The timing, energy and spatial correlations, as well as the capability of pulse shape discrimination provide powerful methods for the measurement of intrinsic radiopurities. The experimental design, detector performance and event-selection algorithms are described. A total of 359*3 kg-days data from three prototypes of CsI(Tl) crystals were taken at China Jinping underground laboratory where offers a good shielding environment. The contamination levels of internal isotopes from Cs-137, Th-232 and U-238 series, as well as the upper bounds of U-235 series are reported. Identification of the whole alpha8 peaks from U/Th decay chains and derivation of those corresponding quenching factors are achieved.
arXiv comment8 pages, 11 figures, 5 tables
References in corpus (11)
- Measurement of Neutrino-Electron Scattering Cross-Section with a CsI(Tl) Scintillating Crystal Array at the Kuo-Sheng Nuclear Power Reactor
- Semi-empirical calculation of quenching factors for ions in scintillators
- New Limits on Interactions between Weakly Interacting Massive Particles and Nucleons Obtained with CsI(Tl) Crystal Detectors
- New Limits on Neutrino Magnetic Moments from the Kuo-Sheng Reactor Neutrino Experiment
- Limits on light WIMPs from the CDEX-1 experiment with a p-type point-contact germanium detector at the China Jingping Underground Laboratory
- Introduction of the CDEX experiment
- First results on low-mass WIMP from the CDEX-1 experiment at the China Jinping underground Laboratory
- Limits on light WIMPs with a germanium detector at 177 eVee threshold at the China Jinping Underground Laboratory
- Measurement of the Intrinsic Radiopurity of Cs-137/U-235/U-238/Th-232 in CsI(Tl) Crystal Scintillators
- Pulse Shape Discrimination Techniques in Scintillating CsI(Tl) Crystals
- Prospects of Scintillating Crystal Detector in Low-Energy Low-Background Experiments