Surface Sensitive Microfluidic Optomechanical Sensors
arXiv:1407.1915 · doi:10.1063/1.4901067
Abstract
The microfluidic optomechanical resonator (uFOMR) based on a thin-walled glass capillary supports high Q-factor (>1000) mechanical modes in the presence of liquids. In this Letter, the sensitivity of the uFOMR to the surface change is studied by layer-by-layer removal of SiO2 molecules from the uFOMR inner surface using various concentrations of hydrofluoric acid solutions. A frequency downshift is observed with a sensitivity of 1.2 Hz/(pg/mm2), which translates to a surface density detection limit of 83 pg/mm2. This work opens a door to using the optomechanical mode for detection and characterization of molecules present near the resonator surface.