Absolute calibration of photon-number-resolving detectors with an analog output using twin beams
arXiv:1401.6071 · doi:10.1063/1.4863433
Abstract
A method for absolute calibration of a photon-number resolving detector producing analog signals as the output is developed using a twin beam. The method gives both analog-to-digital conversion parameters and quantum detection efficiency for the photon fields. Characteristics of the used twin beam are also obtained. A simplified variant of the method applicable to fields with high signal to noise ratios and suitable for more intense twin beams is suggested.
arXiv comment4 pages, 3 figures
References in corpus (13)
- Photon number resolution using a time-multiplexed single-photon detector
- Multiple-photon resolving fiber-loop detector
- Direct measurement and reconstruction of nonclassical features of twin beams generated in spontaneous parametric down-conversion
- Quantum state measurements using multi-pixel photon detectors
- Photon-number statistics with Silicon photomultipliers
- Detection of multimode spatial correlation in PDC and application to the absolute calibration of a CCD camera
- Photon-number distributions of twin beams generated in spontaneous parametric down-conversion and measured by an intensified CCD camera
- Self-consistent characterization of light statistics
- Experimental multi-photon-resolving detector using a single avalanche photodiode
- Self consistent, absolute calibration technique for photon number resolving detectors
- Absolute detector calibration using twin beams
- State reconstruction of a multimode twin beam using photodetection
- Photon-number correlations by photon-number resolving detectors
Cited by in corpus (12)
- Realisation of the first sub shot noise wide field microscope
- Photon number correlation for quantum enhanced imaging and sensing
- Tomography of photon-number resolving continuous-output detectors
- Absolute calibration of an EMCCD camera by quantum correlation linking photon counting to analog regime
- Absolute calibration of a CCD camera with twin beams
- Absolute self-calibration of single-photon and multiplexed photon-number-resolving detectors
- Direct calibration of click-counting detectors
- Quantum differential ghost microscopy
- Microchannel plate cross-talk mitigation for spatial autocorrelation measurements
- Experimental identification of non-classicality of noisy twin beams and other related two-mode states
- On the application of a Silicon photomultiplier-based receiver for binary phase-shift-keying protocols
- Efficient Characterization of N-Beam Gaussian Fields Through Photon-Number Measurements: Quantum Universal Invariants