Herschel SPIRE FTS Relative Spectral Response Calibration
arXiv:1401.2049 · doi:10.1007/s10686-013-9364-z
Abstract
Herschel/SPIRE Fourier transform spectrometer (FTS) observations contain emission from both the Herschel Telescope and the SPIRE Instrument itself, both of which are typically orders of magnitude greater than the emission from the astronomical source, and must be removed in order to recover the source spectrum. The effects of the Herschel Telescope and the SPIRE Instrument are removed during data reduction using relative spectral response calibration curves and emission models. We present the evolution of the methods used to derive the relative spectral response calibration curves for the SPIRE FTS. The relationship between the calibration curves and the ultimate sensitivity of calibrated SPIRE FTS data is discussed and the results from the derivation methods are compared. These comparisons show that the latest derivation methods result in calibration curves that impart a factor of between 2 and 100 less noise to the overall error budget, which results in calibrated spectra for individual observations whose noise is reduced by a factor of 2-3, with a gain in the overall spectral sensitivity of 23% and 21% for the two detector bands, respectively.
15 pages, 13 figures, accepted for publication in Experimental Astronomy
References in corpus (2)
Cited by in corpus (7)
- Far-infrared Herschel SPIRE spectroscopy of lensed starbursts reveals physical conditions of ionised gas
- Systematic characterisation of the Herschel SPIRE Fourier Transform Spectrometer
- The data processing pipeline for the Herschel SPIRE Fourier Transform Spectrometer
- Correcting the extended-source calibration for the Herschel-SPIRE Fourier-Transform Spectrometer
- Herschel SPIRE FTS telescope model correction
- Spectroradiometry with Space Telescopes
- The Herschel SPIRE Fourier Transform Spectrometer Spectral Feature Finder I. The Spectral Feature Finder and Catalogue