Angular and Polarization Response of Multimode Sensors with Resistive-Grid Absorbers
arXiv:1401.1859 · doi:10.1364/JOSAA.31.001557
Abstract
High sensitivity receiver systems with near ideal polarization sensitivity are highly desirable for development of millimeter and sub-millimeter radio astronomy. Multimoded bolometers provide a unique solution to achieve such sensitivity, for which hundreds of single-mode sensors would otherwise be required. The primary concern in employing such multimoded sensors for polarimetery is the control of the polarization systematics. In this paper, we examine the angular- and polarization- dependent absorption pattern of a thin resistive grid or membrane, which models an absorber used for a multimoded bolometer. The result shows that a freestanding thin resistive absorber with a surface resistivity of η/2, where η is the impedance of free space, attains a beam pattern with equal E- and H-plane responses, leading to zero cross polarization. For a resistive-grid absorber, the condition is met when a pair of grids is positioned orthogonal to each other and both have a resistivity of η/2. When a reflective backshort termination is employed to improve absorption efficiency, the cross-polar level can be suppressed below -30 dB if acceptance angle of the sensor is limited to <60degrees. The small cross-polar systematics have even-parity patterns and do not contaminate the measurements of odd-parity polarization patterns, for which many of recent instruments for cosmic microwave background are designed. Underlying symmetry that suppresses these cross-polar systematics is discussed in detail. The estimates and formalism provided in this paper offer key tools in the design consideration of the instruments using the multimoded polarimeters.
arXiv comment22 pages, 15 figures
References in corpus (20)
- Signature of Gravity Waves in Polarization of the Microwave Background
- A Probe of Primordial Gravity Waves and Vorticity
- The Primordial Inflation Explorer (PIXIE): A Nulling Polarimeter for Cosmic Microwave Background Observations
- Planck Early Results: The Planck mission
- Planck Early Results: All sky temperature and dust optical depth from Planck and IRAS: Constraints on the "dark gas" in our galaxy
- ACTPol: A polarization-sensitive receiver for the Atacama Cosmology Telescope
- Benchmark Parameters for CMB Polarization Experiments
- Design, Implementation and Testing of the MAP Radiometers
- CMB Polarization Systematics Due to Beam Asymmetry: Impact on Inflationary Science
- The Relative Performance of Filled and Feedhorn-Coupled Focal-plane Architectures
- The EBEX Experiment
- Characterization of the BICEP Telescope for High-Precision Cosmic Microwave Background Polarimetry
- QUaD: A High-Resolution Cosmic Microwave Background Polarimeter
- The QUIET Instrument
- The Cosmology Large Angular Scale Surveyor (CLASS): 40 GHz optical design
- Properties of a Variable-delay Polarization Modulator
- SPIDER: A Balloon-borne Large-scale CMB Polarimeter
- Instrumental and Analytic Methods for Bolometric Polarimetry
- Antenna-coupled TES bolometers for the Keck Array, Spider, and Polar-1
- Diffraction Considerations for Planar Detectors in the Few-Mode Limit