Measurement of a 127 keV Neutron Field with a micro-TPC Spectrometer
arXiv:1310.6837 · doi:10.1109/TNS.2014.2307093
Abstract
In order to measure the energy of neutron fields, with energy ranging from 8 keV to 1 MeV, a new primary standard is being developed at the IRSN (Institute for Radioprotection and Nuclear Safety). This project, micro-TPC (Micro Time Projection Chamber), carried out in collaboration with the LPSC (Laboratoire de Physique Subatomique et de Cosmologie), is based on the nuclear recoil detector principle. The instrument is presented with the associated method to measure the neutron energy. This article emphasizes the proton energy calibration procedure and energy measurements of a neutron field produced at 127 keV with the IRSN facility AMANDE.
arXiv comment7 pages. Submitted in IEEE TNS/ Special Issue ANIMMA 2013
References in corpus (6)
- Micromegas in a Bulk
- Assessing the discovery potential of directional detection of Dark Matter
- Development of a front end ASIC for Dark Matter directional detection with MIMAC
- Quenching factor measurement in low pressure gas detector for directional dark matter search
- Data acquisition electronics and reconstruction software for real time 3D track reconstruction within the MIMAC project
- A μ-TPC detector for the characterization of low energy neutron fields