Attenuation length measurements of liquid scintillator with LabVIEW and reliability evaluation of the device
arXiv:1305.1471 · doi:10.1088/1674-1137/37/7/076001
Abstract
The attenuation length measuring device was constructed by using oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested with a variety of ways, the test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.10 plus or minus 0.35 m where Gd-LS was heavily used in Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment.
5 pages, 8 figures, 9 conferences
References in corpus (1)
Cited by in corpus (11)
- Production of Gadolinium-loaded Liquid Scintillator for the Daya Bay Reactor Neutrino Experiment
- Slow Liquid Scintillator Candidates for MeV-scale Neutrino Experiments
- A complete optical model for liquid-scintillator detectors
- Rayleigh scattering of linear alkylbenzene in large liquid scintillator detectors
- Rayleigh scattering and depolarization ratio in linear alkylbenzene
- Spectroscopic study of light scattering in linear alkylbenzene for liquid scintillator neutrino detectors
- Measurements of Rayleigh Ratios in Linear Alkylbenzene
- Performance of plastic scintillator modules for top veto tracker at Taishan Antineutrino Observatory
- In-Situ Measurement of Relative Attenuation Length of Gadolinium-Loaded Liquid Scintillator Using Source Data at RENO Experiment
- Determine Energy Nonlinearity and Resolution of and in Liquid Scintillator Detectors by A Universal Energy Response Model
- A novel method for measuring the attenuation length and the group velocity of transparent liquids in a variable length cavity