Improved sensitivity of magnetic measurements under high pressure in miniature ceramic anvil cell for a commercial SQUID magnetometer
arXiv:1304.5301 · doi:10.1063/1.4802832
Abstract
Two modifications have been made to a miniature ceramic anvil high pressure cell (mCAC) designed for magnetic measurements at pressures up to 12.6 GPa in a commercial superconducting quantum interference (SQUID) magnetometer [N. Tateiwa et al., Rev. Sci. Instrum. 82, 053906 (2011)., ibid. 83, 053906 (2012)]. Replacing the Cu-Be piston in the former mCAC with a composite piston composed of the Cu-Be and ceramic cylinders reduces the background magnetization significantly smaller at low temperatures, enabling more precise magnetic measurements at low temperatures. A second modification to the mCAC is the utilization of a ceramic anvil with a hollow in the center of the culet surface. High pressures up to 5 GPa were generated with the "cupped ceramic anvil" with the culet size of 1.0 mm.
arXiv commentRev. Sci. Instrum. 84, 046105 (2013)
References in corpus (2)
Cited by in corpus (7)
- Breakdown of magnetic order in the pressurized Kitaev iridate -LiIrO
- Pressure-induced ferromagnetism with strong Ising-type anisotropy in YbCuSi
- Strong correlation between ferromagnetic superconductivity and pressure-enhanced ferromagnetic fluctuations in UGe
- Superconducting properties of KNaFeAs under pressure
- The structure of the superconducting high-pressure phase of ScCoC
- Design of opposed-anvil-type high-pressure cell for precision magnetometry and its application to quantum magnetism
- Tuning Magnetic and Electronic Properties of Double Perovskite LaCoIrTiO