A measurement of the energy and timing resolution of GlueX Forward Calorimeter using an electron beam
arXiv:1304.4999 · doi:10.1016/j.nima.2013.05.109
Abstract
The performance of the GlueX Forward Calorimeter was studied using a small version of the detector and a variable energy electron beam derived from the Hall B tagger at Jefferson Lab. For electron energies from 110 MeV to 260 MeV, which are near the lower-limits of the design sensitivity, the fractional energy resolution was measured to range from 20% to 14%, which meets the design goals. The use of custom 250 MHz flash ADCs for readout allowed precise measurements of signal arrival times. The detector achieved timing resolutions of 0.38 ns for a single 100 mV pulse, which will allow timing discrimination of photon beam bunches and out-of-time background during the operation of the GlueX detector.
arXiv comment8 pages, 11 figures, submitted to NIMA
References in corpus (1)
Cited by in corpus (12)
- The GlueX Beamline and Detector
- Measurement of the beam asymmetry for and photoproduction on the proton at GeV
- First Results from The GlueX Experiment
- Construction and Performance of the Barrel Electromagnetic Calorimeter for the GlueX Experiment
- Beam Asymmetry for the Photoproduction of and Mesons at GeV
- Search for photoproduction of axion-like particles at GlueX
- Measurement of beam asymmetry for photoproduction on the proton at =8.5 GeV
- Search for axion-like particles through nuclear Primakoff production using the GlueX detector
- Measurement of the Photon Beam Asymmetry in at GeV
- Using machine learning to separate hadronic and electromagnetic interactions in the GlueX forward calorimeter
- First Measurement of Near- and Sub-Threshold Photoproduction off Nuclei
- Searches for Exotic Hadrons at GlueX