paper

Mid-infrared VIPA Spectrometer for Rapid and Broadband Trace Gas Detection

arXiv:1206.1316 · doi:10.1364/OL.37.003285

Abstract

We present and characterize a 2-D imaging spectrometer based on a virtually-imaged phased array (VIPA) disperser for rapid, high-resolution molecular detection using mid-infrared (MIR) frequency combs at 3.1 and 3.8 μm. We demonstrate detection of CH4 at 3.1 μm with >3750 resolution elements spanning >80 nm with ~600 MHz resolution in a <10 μs acquisition time. In addition to broadband detection, rapid, time-resolved single-image detection is demonstrated by capturing dynamic concentration changes of CH4 at a rate of ~375 frames per second. Changes in absorption above the noise floor of 5\times 10-4 are readily detected on the millisecond time scale, leading to important future applications such as real time monitoring of trace gas concentrations and detection of reactive intermediates.