Code-division multiplexing for x-ray microcalorimeters
arXiv:1201.6289 · doi:10.1063/1.3684807
Abstract
We demonstrate the code-division multiplexed (CDM) readout of eight transition-edge sensor microcalorimeters. The energy resolution is 3.0 eV (full width at half-maximum) or better at 5.9 keV, with a best resolution of 2.3 eV and a mean of 2.6 eV over the seven modulated detectors. The flux-summing CDM system is described and compared with similar time-division multiplexed (TDM) readout. We show that the sqrt(Npixels) multiplexing disadvantage associated with TDM is not present in CDM. This demonstration establishes CDM as both a simple route to higher performance in existing TDM microcalorimetric experiments and a long-term approach to reaching higher multiplexing factors.
To be published in Applied Physics Letters
References in corpus (1)
Cited by in corpus (8)
- HOLMES - The Electron Capture Decay of 163Ho to Measure the Electron Neutrino Mass with sub-eV sensitivity
- A Review of X-ray Microcalorimeters Based on Superconducting Transition Edge Sensors for Astrophysics and Particle Physics
- High-resolution gamma-ray spectroscopy with a microwave-multiplexed transition-edge sensor array
- The use of low temperature detectors for direct measurements of the mass of the electron neutrino
- Optimising the multiplex factor of the frequency domain multiplexed readout of the TES-based microcalorimeter imaging array for the X-IFU instrument on the Athena Xray observatory
- On-sky performance of the SPT-3G frequency-domain multiplexed readout
- Frequency Shift Algorithm: Design of a Baseband Phase Locked Loop for Frequency-Domain Multiplexing Readout of X-ray Transition-Edge Sensor Microcalorimeters
- Development of microwave superconducting microresonators for neutrino mass measurement in the HOLMES framework