Sensitivity of synthetic aperture laser optical feedback imaging
arXiv:1201.3282 · doi:10.1364/JOSAA.29.000476
Abstract
In this paper we compare the sensitivity of two imaging configurations both based on Laser Optical Feedback Imaging (LOFI). The first one is direct imaging, which uses conventional optical focalisation on target and the second one is made by Synthetic Aperture (SA) Laser, which uses numerical focalisation. We show that SA configuration allows to obtain good resolutions with high working distance and that the drawback of SA imagery is that it has a worse photometric balance in comparison to conventional microscope. This drawback is partially compensated by the important sensitivity of LOFI. Another interest of SA relies on the capacity of getting a 3D information in a single x-y scan.
References in corpus (4)
- Measuring the Transmission Matrix in Optics : An Approach to the Study and Control of Light Propagation in Disordered Media
- Comparative study of autodyne and heterodyne laser interferometry for imaging
- Experimental comparison of autodyne and heterodyne laser interferometry using a Nd:YVO4 microchip laser
- Self-aligned setup for laser optical feedback imaging insensitive to parasitic optical feedback
Cited by in corpus (4)
- Acousto-optic laser optical feedback imaging
- Synthetic aperture laser optical feedback imaging using a translational scanning with galvanometric mirrors
- Deep and optically resolved imaging through scattering media by space-reversed propagation
- Limitations of synthetic aperture laser optical feedback imaging