Characterization of a broad energy germanium detector and application to neutrinoless double beta decay search in Ge-76
arXiv:1012.5200 · doi:10.1088/1748-0221/6/04/P04005
Abstract
The performance of a 630g commercial broad energy germanium (BEGe) detector has been systematically investigated. Energy resolution, linearity, stability vs. high-voltage (HV) bias, thickness and uniformity of dead layers have been measured and found to be excellent. Special attention has been dedicated to the study of the detector response as a function of bias HV. The nominal depletion voltage being 3000V, the detector under investigation shows a peculiar behavior for biases around 2000V: in a narrow range of about 100V the charge collection is strongly reduced. The detector seems to be composed by two parts: a small volume around the HV contact where charges are efficiently collected as at higher voltage, and a large volume where charges are poorly collected. A qualitative explanation of this behavior is presented. An event-by-event pulse shape analysis based on A/E (maximum amplitude of the current pulse over the total energy released in the detector) has been applied to events in different energy regions and found very effective in rejecting non localized events. In conclusion, BEGe detectors are excellent candidates for the second phase of GERDA, an experiment devoted to neutrinoless double beta decay of Ge-76.
arXiv commentSubmitted to JINST, 17 pages, 11 figures. Companion paper of arXiv:1012.4300
References in corpus (4)
- \textsc{MaGe} - a {\sc Geant4}-based Monte Carlo Application Framework for Low-background Germanium Experiments
- Pulse shape discrimination studies with a Broad-Energy Germanium detector for signal identification and background suppression in the GERDA double beta decay experiment
- Signal modeling of high-purity Ge detectors with a small read-out electrode and application to neutrinoless double beta decay search in Ge-76
- Cosmic Microwave Background Polarization
Cited by in corpus (8)
- Production, characterization and operation of Ge enriched BEGe detectors in GERDA
- Determining the Drift Time of Charge Carriers in P-Type Point-Contact HPGe Detectors
- GELATIO: a general framework for modular digital analysis of high-purity Ge detector signals
- Characterization of High-Purity Germanium Detectors with Amorphous Germanium Contacts in Cryogenic Liquids
- Isotopically modified Ge detectors for {\sc Gerda}: from production to operation
- Characterization of an electrically cooled BEGe detector till E MeV
- Characterization study of a broad-energy germanium detector at CJPL
- Alpha Backgrounds for HPGe Detectors in Neutrinoless Double-Beta Decay Experiments