Measurement of Reciprocity Failure in Near Infrared Detectors
arXiv:1012.2411 · doi:10.1086/658282
Abstract
Flux dependent non-linearity (reciprocity failure) in HgCdTe near infrared detectors can severely impact an instrument's performance, in particular with respect to precision photometric measurements. The cause of this effect is presently not understood. To investigate reciprocity failure, a dedicated test system was built. For flux levels between 1 and 50,000 photons/s, a sensitivity to reciprocity failure of approximately 0.1%/decade was achieved. A wavelength independent non-linearity due to reciprocity failure of about 0.35%/decade was measured in a 1.7 micron HgCdTe detector.
arXiv comment7 pages, 8 figures
Cited by in corpus (11)
- Characterization of the atmosphere of the hot Jupiter HAT-P-32Ab and the M-dwarf companion HAT-P-32B
- Laboratory measurement of the brighter-fatter effect in an H2RG infrared detector
- Reciprocity Failure in HgCdTe Detectors: Measurements and Mitigation
- The Cosmic Infrared Background Experiment (CIBER): The Wide-Field Imagers
- A Synthetic Roman Space Telescope High-Latitude Imaging Survey: Simulation Suite and the Impact of Wavefront Errors on Weak Gravitational Lensing
- Point spread function modelling for astronomical telescopes: a review focused on weak gravitational lensing studies
- The Cosmic Infrared Background Experiment (CIBER): The Low Resolution Spectrometer
- The effect of detector nonlinearity on WFIRST PSF profiles for weak gravitational lensing measurements
- The SPHEREx Sky Simulator: Science Data Modeling for the First All-Sky Near-Infrared Spectral Survey
- An Algorithm for Real-Time Optimal Photocurrent Estimation including Transient Detection for Resource-Constrained Imaging Applications
- Impact of Image Persistence in the Roman Space Telescope High-Latitude Survey