Extended Depth of Field for High Resolution Scanning Transmission Electron Microscopy
arXiv:1010.4500 · doi:10.1017/S1431927610094171
Abstract
Aberration-corrected scanning transmission electron microscopes (STEM) provide sub-angstrom lateral resolution; however, the large convergence angle greatly reduces the depth of field. For microscopes with a small depth of field, information outside of the focal plane quickly becomes blurred and less defined. It may not be possible to image some samples entirely in focus. Extended depth-of-field techniques, however, allow a single image, with all areas in-focus, to be extracted from a series of images focused at a range of depths. In recent years, a variety of algorithmic approaches have been employed for bright field optical microscopy. Here, we demonstrate that some established optical microscopy methods can also be applied to extend the ~6 nm depth of focus of a 100 kV 5th-order aberration-corrected STEM (alpha_max = 33 mrad) to image Pt-Co nanoparticles on a thick vulcanized carbon support. These techniques allow us to automatically obtain a single image with all the particles in focus as well as a complimentary topography map.
Accepted, Microscopy and Microanalysis
Cited by in corpus (5)
- Nanomaterial datasets to advance tomography in scanning transmission electron microscopy
- Breaking the Crowther Limit: Combining Depth-Sectioning and Tilt Tomography for High-Resolution, Wide-Field 3D Reconstructions
- Quantum Wavefront Shaping with a 48-element Programmable Phase Plate for Electrons
- Electron tomography for functional nanomaterials
- The Limits of Resolution and Dose for Aberration-Corrected Electron Tomography