Abstract
We report on the performance of ultra low noise differential photodetectors especially designed for probing of atomic ensembles with weak light pulses. The working principle of the detectors is described together with the analysis procedures employed to extract the photon shot noise of light pulses with s duration. As opposed to frequency response peaked detectors, our approach allows for broadband quantum noise measurements. The equivalent noise charge (ENC) for two different hardware approaches is evaluated to 280 and 340 electrons per pulse, respectively which corresponds to a dark noise equivalent photon number of and in the two approaches. Finally, we discuss the possibility of removing classical correlations in the output signal caused by detector imperfection by using double--correlated sampling methods.
7 pages, 8 figures, accepted for publication in Meas. Sci. Technol
References in corpus (5)
- Mesoscopic atomic entanglement for precision measurements beyond the standard quantum limit
- Non-Destructive Probing of Rabi Oscillations on the Cesium Clock Transition near the Standard Quantum Limit
- Conditions for spin squeezing in a cold 87Rb ensemble
- Rayleigh superradiance and dynamic Bragg gratings in an end-pumped Bose-Einstein condensate
- Observation of Topological and Parity-dependent Phase of Spin States
Cited by in corpus (6)
- Interaction-based quantum metrology showing scaling beyond the Heisenberg limit
- Real-time vector field tracking with a cold-atom magnetometer
- Quantum atom-light interfaces in the gaussian description for spin-1 systems
- Heterodyne non-demolition measurements on cold atomic samples: towards the preparation of non-classical states for atom interferometry
- Real-time shot-noise-limited differential photodetection for atomic quantum control
- Optimal Signal Recovery for Pulsed Balanced Detection