particle physics instrumentation

Summary of quality control (QC) of ATLAS18 production ITk strip sensors

arXiv:2607.14691 · doi:10.1016/j.nima.2026.171716

summary

The paper describes the quality‑control program for the ATLAS ITk silicon strip sensors, detailing standardized test procedures, data handling tools, and production‑yield results for the HL‑LHC upgrade.

Abstract

To address the demanding operational requirements of the High-Luminosity upgrade of the Large Hadron Collider (HL-LHC), the ATLAS experiment is replacing its current Inner Detector with a new all-silicon Inner Tracker (ITk). The ITk will feature an active area of 165 m2, with its outer tracking layers populated by approximately 18,000 ATLAS18 n+-in-p silicon strip sensors. The silicon sensors, available in eight geometries tailored to two barrel and six endcap types, respectively, are designed to tolerate fluences of up to 1.6 x 10^15 neq/cm2 and ionizing doses of 66 Mrad. A comprehensive, multi-year Quality Control (QC) program is underway across multiple international institutes to evaluate these ITk strip sensors for mechanical and electrical conformity. The QC process includes IV/CV characterization, full strip tests, long-term current stability monitoring, visual inspection, and metrology tests. To manage the high throughput of about 500 sensors per month, the collaboration has implemented standardized test procedures, software packages for data monitoring and integrity checks, unified data formats, and automated analysis tools. The standardization ensures consistent pass/fail evaluation and centralized data handling that enables effective identification of trends and anomalies at all sites during the production. This contribution presents an overview of the ITk strip sensor production and QC framework, along with key findings throughout the whole production, such as charge-up of sensors, stability of the leakage currents, nonrecoverable IV breakdown, and low inter-strip isolation within wafers. It provides insights into sensor yield, quality trends, and reviews specific case studies, such as p-stop doping non-uniformity. Over 91% of the production, totaling over 590 batches, were tested and accepted. Six batches were rejected. These account for 2.8% of the total tested sensors.

7 pages

Topics & keywords

#silicon strip sensors#quality control#ATLAS inner tracker#HL-LHC detector production#automated testingn+-in-p siliconIV/CV characterizationleakage current stabilityp-stop doping uniformitysensor yieldautomated analysis tools