A Novel Low-Background Photomultiplier Tube Developed for Xenon Based Detectors
arXiv:2412.10830 · doi:10.1016/j.nima.2025.170290
Abstract
Photomultiplier tubes (PMTs) are essential in xenon detectors like PandaX, LZ, and XENON experiments for dark matter searches and neutrino properties measurement. To minimize PMT-induced backgrounds, stringent requirements on PMT radioactivity are crucial. A novel 2-inch low-background R12699 PMT has been developed through a collaboration between the PandaX team and Hamamatsu Photonics K.K. corporation. Radioactivity measurements conducted with a high-purity germanium detector show levels of approximately 0.08 mBq/PMT for and 0.06~mBq/PMT for the late chain, achieving a 15-fold reduction compared to R11410 PMT used in PandaX-4T. The radon emanation rate is below 3.2 Bq/PMT (@90\% confidence level), while the surface activity is less than 18.4 Bq/cm. The electrical performance of these PMTs at cryogenic temperature was evaluated. With an optimized readout base, the gain was enhanced by 30\%, achieving an average gain of at -1000~V and -100~C. The dark count rate averaged 2.5~Hz per channel. Compactness, low radioactivity, and robust electrical performance in the cryogenic temperature make the R12699 PMT ideal for next-generation liquid xenon detectors and other rare event searches.