PET monitoring of cancer therapy with He-3 and C-12 beams: a study with the GEANT4 toolkit

arXiv:0708.1691 · doi:10.1088/0031-9155/52/24/007

Abstract

We study the spatial distributions of $β^+$-activity produced by therapeutic beams of He and C ions in various tissue-like materials. The calculations were performed within a Monte Carlo model for Heavy-Ion Therapy (MCHIT) based on the GEANT4 toolkit. The contributions from C, N, O, F and P positron-emitting nuclei were calculated and compared with experimental data obtained during and after irradiation. Positron emitting nuclei are created by C beam in fragmentation reactions of projectile and target nuclei. This leads to a $β^+$-activity profile characterised by a noticeable peak located close to the Bragg peak in the corresponding depth-dose distribution. On the contrary, as the most of positron-emitting nuclei are produced by He beam in target fragmentation reactions, the calculated total $β^+$-activity during or soon after the irradiation period is evenly distributed within the projectile range. However, we predict also the presence of N, O, F created in charge-transfer reactions by low-energy He ions close to the end of their range in several tissue-like media. The time evolution of $β^+$-activity profiles was investigated for both kinds of beams. Due to the production of F nuclide the $β^+$-activity profile measured 2 or 3 hours after irradiation with He ions will have a distinct peak correlated with the maximum of depth-dose distribution. We found certain advantages of low-energy He beams over low-energy proton beams for reliable PET monitoring during particle therapy of shallow located tumours. In this case the distal edge of $β^+$-activity distribution from F nuclei clearly marks the range of He in tissues.